1. Low recovery rate
Possible reasons and solutions
1)Cartridge activation conditions are not appropriate – correct activation of SPE cartridges depending on the stationary phase.
2)The solvent of the sample has a stronger force on the target component than the stationary phase-select the SPE cartridge with more selectivity to the target component; adjusting the pH of the sample solvent to increase the force of the target component on the stationary phase; changing the polarity of the sample solvent, reducing the force of the target ingredient in the solvent.
3)Improper cleaning solvent selection, too strong elution ability – use the correct cleaning solvent, choose the solvent with weaker elution ability.
4)The flow rate is too fast when loading – gravity natural load or control sample flow rate ≤ 1 ml/min.
5)SPE cartridge is too small – replace the larger size SPE cartridge; use a more selective SPE cartridge; use a larger sample of SPE cartridge.
6)SPE cartridge cleaning solvent is not sufficiently dried before elution – fully drain the rinse solvent.
7)The flow rate is too fast or too slow when eluting – control flow rate 1-2ml / min.
Possible reasons and solutions
1)Cartridge activation conditions are not appropriate – correct activation of SPE cartridges depending on the stationary phase.
2)The solvent of the sample has a stronger force on the target component than the stationary phase-select the SPE cartridge with more selectivity to the target component; adjusting the pH of the sample solvent to increase the force of the target component on the stationary phase; changing the polarity of the sample solvent, reducing the force of the target ingredient in the solvent.
3)Improper cleaning solvent selection, too strong elution ability – use the correct cleaning solvent, choose the solvent with weaker elution ability.
4)The flow rate is too fast when loading – gravity natural load or control sample flow rate ≤ 1 ml/min.
5)SPE cartridge is too small – replace the larger size SPE cartridge; use a more selective SPE cartridge; use a larger sample of SPE cartridge.
6)SPE cartridge cleaning solvent is not sufficiently dried before elution – fully drain the rinse solvent.
7)The flow rate is too fast or too slow when eluting – control flow rate 1-2ml / min.
2. The target component cannot be eluted from the SPE cartridge.
Possible reasons and solutions
1)The selectivity relative to the target component is too strong-select the cartridge that retains the weaker analyte; select the elution solvent with stronger elution ability.
2) SPE cartridge size is too large – increase the amount of eluent; replace it with a smaller specification SPE cartridge.
3) Insufficient amount of eluent – increase eluent.
4) The flow rate is too fast or too slow when eluting – control flow rate 1-2ml / min.
5) The elution ability is too weak – adjust the pH of the eluent or change the polarity of the eluent to increase the elution ability of the solvent to the target component.
Possible reasons and solutions
1)The selectivity relative to the target component is too strong-select the cartridge that retains the weaker analyte; select the elution solvent with stronger elution ability.
2) SPE cartridge size is too large – increase the amount of eluent; replace it with a smaller specification SPE cartridge.
3) Insufficient amount of eluent – increase eluent.
4) The flow rate is too fast or too slow when eluting – control flow rate 1-2ml / min.
5) The elution ability is too weak – adjust the pH of the eluent or change the polarity of the eluent to increase the elution ability of the solvent to the target component.