Solutions For High Pressure And Baseline Drift In HPLC Columns

This is the most common problem in the use of high performance liquid phase, which refers to the sudden rise of pressure, generally due to the blockage in the flow path. At this time, we should check by sections.
Disconnect the inlet of the vacuum pump.
At this time, the peek tube is filled with liquid, so that the peek tube is lower than the solvent bottle, to see whether the liquid drops freely. If the liquid does not drop or drops slowly, the solvent filter head is blocked. Treatment method: soak in 30% nitric acid for half an hour, then rinse with ultra pure water. If the liquid drips freely and the solvent filter head is normal, check again;
Open the purge valve
If the pressure does not drop obviously, it will be blocked by filter white head. Treatment method: take out the filtered white head and use 10% isopropanol for ultrasonic for half an hour. If the pressure drops below 100psi (6.7bar), the filter white head is normal, and recheck;
Remove the column outlet end
If the pressure does not drop, the HPLC column is blocked. Treatment: if the buffer salt is blocked, flush 95% water to normal pressure. If some strongly retained substances cause blockage, the mobile phase stronger than the current one should be used to rush to normal pressure. If the washing pressure does not drop for a long time according to the above method, the inlet and outlet of the column can be connected to the instrument in turn, and the column can be washed with mobile phase. At this time, if the column pressure still does not drop, only change the column inlet sieve plate, but once the operation is not very good, it is easy to cause column efficiency drop, so try to useless.

没有评论:

发表评论