Use of pipette:
a. Set liquid volume:
1). When adjusting from large volume to small volume, you need to use the normal
adjustment method and rotate the scale anticlockwise.
2). When adjusting from small volume to large volume, adjust clockwise to the scale
exceeding the set volume and then adjust back to the set volume to ensure the best accuracy.
a. Set liquid volume:
1). When adjusting from large volume to small volume, you need to use the normal
adjustment method and rotate the scale anticlockwise.
2). When adjusting from small volume to large volume, adjust clockwise to the scale
exceeding the set volume and then adjust back to the set volume to ensure the best accuracy.
b. Assemble pipette suction head:
1). Insert the pipette end into the suction head vertically with a single channel, rotate
it slightly from left to right and tighten it.
2). Prepare multi-channel pipette, align the first line of the pipette with the first head, insert it at an Oblique Angle, shake it a little bit to tighten it, and slightly exceed the o-ring after the head is inserted.
1). Insert the pipette end into the suction head vertically with a single channel, rotate
it slightly from left to right and tighten it.
2). Prepare multi-channel pipette, align the first line of the pipette with the first head, insert it at an Oblique Angle, shake it a little bit to tighten it, and slightly exceed the o-ring after the head is inserted.
Curing of Pipette
1). If the liquid accidentally enters the piston chamber, remove the pollution in time;
2). After using the pipette, you need to adjust the range of the pipette to the maximum value, and place the pipette vertically on the pipette rack;
3). According to the frequent use, all pipettes should be regularly washed with soapy water or with 60% isopropanol disinfection, then washed with double steam water and dried;
4). Avoid putting in at a higher temperature to prevent deformation and leakage or inaccurate;
5). Reach out to professionals staff to deal with problems in time;
6). Do not place the pipette horizontally or upside down when there is liquid in the pipette suction nozzle, so as to prevent the liquid from flowing into the piston chamber to corrode the pipette piston;
2). After using the pipette, you need to adjust the range of the pipette to the maximum value, and place the pipette vertically on the pipette rack;
3). According to the frequent use, all pipettes should be regularly washed with soapy water or with 60% isopropanol disinfection, then washed with double steam water and dried;
4). Avoid putting in at a higher temperature to prevent deformation and leakage or inaccurate;
5). Reach out to professionals staff to deal with problems in time;
6). Do not place the pipette horizontally or upside down when there is liquid in the pipette suction nozzle, so as to prevent the liquid from flowing into the piston chamber to corrode the pipette piston;